Total Hydrocarbon Analyzer
4000 Series Overview:
All 4000 series units employ the same basic architecture and electronics; however, each analyzer has additional features:
4020 Total Hydrocarbon Analyzer
Uses FID to measure total hydrocarbons in ppm in a sample gas stream
4030 Total Hydrocarbon Analyzer
Basic 4020 with sample system at a higher temperature for heated applications
Avoids condensation of the sample
4040 Methane / Non-Methane
Basic 4020, plus temperature-controlled sampling system with GC column to give accurate readings of both methane and non-methane hydrocarbons in a gas stream
4060 Benzene Analyzer
Same as 4040, but used to measure ppb levels of benzene instead of methane/non-methane
4080 Total Hydrocarbon Analyzer
Basic 4020 with sample system including stripper to measure total hydrocarbons in an aqueous sample
• Easy to use software, user programmable ranges
• GC technology used for complete separation of Methane, Ethylene Oxide, Benzene, Styrene, BTX and many other compounds (application dependent)
• Automatic fuel shut off system
• Electronic flame-out guard circuit
• Adjustable alarm and oven settings
• Fast response
• Sample filter with exchangeable stainless steel element
• Built-in sample pump (optional)
• Stainless steel sample system with flexible design to handle either a 100% H2, H2/N2 or H2/He support gas fuel mixture
• Automatic ignition
• X-purged configuration for hazardous area installations available
• Standard auto-calibration programming, ideal for unmanned plant conditions
• Enhanced temperature control of sample compartment for stable, reproducible results
• Easily accessible pressure and flow control devices
• Slide out chassis design for easy access to internal components
• CE marked
• Optional methanizer for CO and CO2 measurements
Ranges | 0-10, 0-100, 0-1000 ppm THC. Other ranges available |
Sensitivity | 1% of FS |
Resolution | 0.1 ppm |
Accuracy | ± 2% of FS at constant temperature |
Response Time | 90% in less than 15 seconds, data updated every 5 minutes |
Temperature | 40° to 110° (4.44° to 43.33°C) |
Fuel | UHP Hydrogen @ 30 PSI incoming pressure (standard) |
Power | 115 or 220 VAC, 50/60Hz, 600VA |
Noise | Less than 0.5% of FS |
Drift | 2% FS per week |
Linearity | 1% FS through all ranges |
Repeatability | 1% FS through all ranges |
Sample Chamber | 212° to 248°F (100° to 120°C) per application |
Flow Rate | 1 Liter/minute (standard) |
Outputs | 0-1 VDC & 4-20 mADC, isolated current output |
Warm-Up Time | Usable in approximately 4 hours |
Carrier Gas | UHP Nitrogen @ 50 PSI incoming pressure |
Enclosure | 19″ rack mounted steel enclosure |
• Ambient monitoring of THC
• THC measurement of automotive exhaust