The 3300 series of low cost, microprocessor-based instrument are specially designed to satisfy industry’s need for a simple yet highly reliable process analyzer. These units offer excellent accuracy, ease of use, high reliability, low maintenance and a variety of standard features usually found on more expensive instruments.
Multiple Configurations
This series is a cost-effective solution to many process applications. With three different sensing technologies to choose from and multiple design configurations, a customized unit for a given application is easily secured. Sensing ranges are varied based on the oxygen sensor and technology used. All units come with an auto-ranging capability and a 0-25% calibration range. Fully adjustable high or low concentration alarms provide the versatility to satisfy nearly any requirement.
Outputs for Data Transmission
A standard 0-10 VDC output provides range identification, while the oxygen concentration is output through 4-20 mADC negative ground and 0-10 VDC negative ground signals. An optional unidirectional RS-232 serial interface can be incorporated to relay information to a host computer for remote monitoring, making information and analysis results as close as a personal computer.
Sensing Technologies
The 3300 Series is available with three sensing technologies, each with their own benefit.
Micro-fuel Cell – 3300T, 3300P
• Disposable, no maintenance sensor design
• Specific to oxygen
• Linear from 0-100% O2
• Absolute zero (no O2, no output); single point calibration only
• Unaffected by hydrocarbons, H2 or CO2
• Easy to handle, inexpensive and easy to replace
• Can apply to ppb, ppm and percent ranges
Zirconium Oxide – 3300Z
• Quick response time
• Long life / service in clean, non-corrosive streams
• No shelf life concerns
• Can be applied on a wide variety of applications
Paramagnetic – 3300M
• Long life / service in clean gas streams
• High accuracy
• Can be applied on corrosive, high temperature a
Additional Advantages
• Customer selectable ranges
• Auto ranging capabilities
• Fully adjustable high or low alarms set-points with corresponding relay contacts
• Digital interface allows monitoring from a remote station
• Sensor failure alarm
• Principle: Oxygen specific* electrochemical transducer Micro-fuel Cell Class B-2C
• Ranges: Two ranges, with auto-ranging capability; field configurable between 0-10 ppm and 10,000 ppm O2
• Calibration range: 0-25% O2
• Signal outputs:
• For oxygen concentration: 4-20 mADC negative ground and 0-10 VDC signals
• For range identification: 0-10 VDC signal
• Alarm outputs:
• Two user configurable setpoints with corresponding relay contacts
• Display: 4 digit LED
• Power requirements: Universal AC power supply 100-240 VAC, 47-63 Hz; 10 watts maximum
• Air separation and liquefaction
• Pure, gaseous hydrocarbon stream monitoring
• Emissions monitoring
• Protective atmosphere blanketing of primary liquid feedstocks and flammable liquids
• Process monitoring of gaseous monomers – vinyl chloride, propylene, butadiene, isoprene, or ethylene
• Gas purity certification