Explosion Proof Paramagnetic Oxygen Analyzer​​​​

The Model 3020M is a versatile microprocessor-based instrument specifically designed to detect oxygen in a variety of gases. It is a precise, cost effective instrument capable of measuring percent oxygen in as low as the 0-1% range. Drawing on the solid performance of the 3000 series control platform, the 3020M provides the user with a wide variety of features and options to satisfy virtually any application.

Paramagnetism

The physical property distinguishing oxygen from other gases is its paramagnetism, which is significantly greater than other common gases. Consequently, the molecules of oxygen are attracted by strong magnetic fields which can be used in detection. The sensor consists of two spheres arranged in the form of a dumbbell suspended in a symmetrical, non-uniform magnetic field. When the surrounding gas contains oxygen, the spheres rotate out of the magnetic field by the relatively strong attraction of oxygen. A light beam focused on a mirror attached reflects asymmetrically onto two photo diodes resulting in a voltage shift. The difference of this produces a current used to drive the dumbbells back to the original position. The current flow required to maintain the null position is directly proportional to the oxygen concentration and is shown linearly in volume percent oxygen in the display. Properly maintained sensors last for years with little or no attention making them ideal for critical measurements in the low percent range.

Flexibility

This unit offers three dynamic, user-configurable ranges plus auto ranging and can be calibrated in any range. It is linear on all three ranges, eliminating the need to recalibrate while switching between ranges. This analyzer comes standard with an isolated 4-20 mADC output for oxygen concentration and range identification. Additionally, a bi-directional RS-232C serial communication interface provides for remote monitoring and control of span and zero functions.

Additional Advantages

• Linearity of analysis across three user-selectable ranges

• Auto-ranging to follow process upsets

• Auto-calibration electronics standard

• No purge gas requirements

Options

C Electronically operated cal/zero valves

F Flame arrestors for Class 1, Div 1, Group C/D service

G Flame arrestors for Class 1, Div 1, Group C/D service with cal valves

H Flame arrestors for Group B (hydrogen) service

I Flame arrestors for Group B service with cal valves


Sample Gas Conditioning Requirements

• Sample Inlet Pressure: 5 to 20 psig (regulate to fixed inlet pressure)

• Sample Flow: 1 to 2 SCFH

• Sample Dewpoint: Non-condensing at ambient temperature

• Sample Temperature: 32° to 122°F (0° to 50°C)

• Detector Wetted Parts: Glass, steel, gold, Viton, acrylic glass and epoxy resin